使用Strelow 法测定不同浓度盐酸中铁和镉在DOWEX 1× X8阴离子交换树脂上的分配系数。结果显示,在1 mol/L盐酸中,铁(Ⅲ)的 Kd值低于10 mL/g ,不能吸附在树脂上;而镉(Ⅱ)的 K d值为2281 m L/g ,强吸附在树脂上,因此可以实现两者的有效分离。使用盐酸溶解餐具用钢样品,溶液通过装填了DOWEX 1× X8阴离子交换树脂的小型柱,铁(Ⅲ)可以顺利通过,而镉(Ⅱ)吸附在离子柱中,从而实现镉与铁的有效分离和富集。然后用1 mL 1.6 mol/L硝酸洗下吸附在柱子上的镉,收集的溶液使用电感耦合等离子体原子发射光谱法(ICP‐A ES )测定,从而建立了小型离子交换柱分离‐电感耦合等离子体原子发射光谱法测定餐具用钢中微量镉的方法。镉的质量浓度在0.010~0.10 m g/L范围内与发射强度呈线性,校准曲线的线性相关系数为0.9999。方法中镉的检出限为0.004 g/m L (相当于实际样品中镉的质量分数为0.04μg/g )。按照实验方法测定餐具用钢样品中镉含量,结果的相对标准偏差(RSD ,n=7)为5.8%~7.3%,加标回收率为90%~100%。
The distribution coefficient of iron and cadmium in hydrochloric acid with various concentrations on DOWEX 1 × X8 anion exchange resin was determined by Strelow method .The results showed that the Kd value of iron (Ⅲ) in 1 mol/L hydrochloric acid was less than 10 mL/g ,indicating that it could not be adsorbed on the resin .However ,the Kd value of cadmium (Ⅱ) was 2 281 mL/g ,indicating that it could be strongly adsorbed on the resin .Therefore ,iron (Ⅲ) and cadmium (Ⅱ) could be effectively separated . The steel sample for cutlery was dissolved with hydrochloric acid ,then the sample solution flowed through the small column which was packed with DOWEX 1 × X8 anion exchange resin .The iron (Ⅲ) in solution could pass through the column ,while cadmium (Ⅱ) would be adsorbed onto the column ,realizing the ef‐fective separation and enrichment of cadmium and iron .Then ,the cadmium (Ⅱ ) on column was eluted with 1 mL of 1.6 mol/L nitric acid .The content of cadmium in collected solution was determined by induc‐tively coupled plasma atomic emission spectrometry (ICP‐AES) .Consequently ,the determination method of micro cadmium in steel for cutlery by ICP‐AES after separation with small ion exchange column was es‐tablished .T he mass concentration of cadmium in range of 0. 010‐0. 10 mg/L w as linear to the emission in‐tensity .The linear correlation coefficient of calibration curve was 0.999 9 .The detection limit of cadmium was 0.004 g/mL (which was equivalent to the mass fraction of 0.04 μg/g of cadmium in actual sample) . The content of cadmium in steel for cutlery was determined according to the experimental method .The rel‐ative standard deviation (RSD ,n=7) was 5.8%‐7.3% .The recoveries were between 90% and 100% .
参考文献
[1] | Waalkes MP.Cadmium carcinogenesis.[J].Mutation Research: International Journal on Mutagenesis, Chromosome Breakage and Related Subjects,20031/2(1/2):107-120. |
[2] | 宋传旺;庞晋山.电感耦合等离子体原子发射光谱同时测定不锈钢厨具铬、镍、镉、铅、砷迁移量[J].理化检验-化学分册,2010(5):528-530. |
[3] | 郑国经.电感耦合等离子体原子发射光谱分析仪器与方法的新进展[J].冶金分析,2014(11):1-10. |
[4] | 蔡鹭欣;普旭力;邹建龙;董清木;王鸿辉.ICP-AES法测定氧化铁皮中铅、铬、镉[J].福建分析测试,2011(6):54-56. |
[5] | 周明辉;唐志锟;宋武元;萧达辉.六亚甲基四胺剥离-电感耦合等离子体原子发射光谱法测定钢材镀锌层中铅镉砷[J].冶金分析,2012(1):48-51. |
[6] | Hiroaki Matsumiya;Yuto Sakane;Masataka Hiraide.Determination of trace impurities in high-purity iron using salting-out of polyoxyethylene-type surfactants[J].Analytica chimica acta,20091(1):86-90. |
[7] | Hiroaki Matsumiya;Shigeru Furuzawa;Masataka Hlralde.Removal of an Iron Matrix with Polyoxyethylene-Type Surfactant-Coated Amberlite XAD-4 for the Determination of Trace Impurities in High-Purity Iron[J].Analytical chemistry,200516(16):5344-5348. |
[8] | Serife Tokahoglu;Senol Kartal;Latif Elci.Determination of Some Trace Elements in High-Purity Aluminium, Zinc and Commercial Steel by AAS After Preconcentration on Amberlite XAD-1180 Resin[J].Mikrochimica Acta: An International Journal for Physical and Chemical Methods of Analysis,19973/4(3/4):281-286. |
[9] | YUYU XU;HUI WANG;GUOXIN WANG;CHENG WU;JINFAN ZHOU.Determination of Trace Amounts of Rare-Earth Elements in Clean Steel by Inductively Coupled Plasma Optical Emission Spectrometry after Removal of Iron Matrix with Anion-Exchange Resin[J].Applied Spectroscopy: Society for Applied Spectroscopy,20105(5):543-546. |
[10] | 刘烽;王慧;王国新;许玉宇;张维伟;吴广宇;周锦帆.铁及镉在Dowex 1-X8阴离子交换树脂上分配系数的测定及其用于钢中微量镉的分析[J].化学试剂,2013(5):432-434,454. |
[11] | 谢晓雁;吴文启;李奋;廖红梅;刘媛媛.离子交换分离-电感耦合等离子体原子发射光谱法测定锰矿中铊[J].冶金分析,2015(8):61-65. |
- 下载量()
- 访问量()
- 您的评分:
-
10%
-
20%
-
30%
-
40%
-
50%