为了消除黄金饰品检测中金基体抑制效应和干扰作用,提高测试效率,将钌、钯、铱、铂的单元素标准溶液配制成多元素混合标准溶液,将其加入到配好的高纯金基体溶液中混匀定容配制成含金体系标准溶液系列.采用电感耦合等离子体原子发射光谱法(ICP-AES)测试该混合标准溶液的稳定性,稳定性检验时间为12个月.结果显示,不同日期该混合标准溶液系列的校准曲线回归方程相关系数均大于0.999;t检验结果表明含金体系中钌、钯、铱、铂混合标准溶液在12个月内浓度无显著性变化;15~30℃避光条件下可以稳定储存.
In order to eliminate the inhibitory effect and interference suppression of gold matrix in gold ac-cessories detection as well as to improve the testing efficiency ,the single element standard solution of Ru , Pd ,Ir and Pt should be mixed to prepare multi-element standard solution ,which was then added into high-purity gold matrix solution to prepare standard solution series .The stability of this mixed standard solu-tion was tested by inductively coupled plasma atomic emission spectrometry (ICP-AES) .The testing peri-od was 12 month .The results showed that the correlation coefficients of calibration curves at different sta-ges were higher than 0 .999 .The t test results also indicated that the concentration of Ru ,Pd ,Ir and Pt in gold system had no significant changes within a year .The prepared mixed standard solution could be stably kept for long term in darkness at 15-30 ℃ .
参考文献
[1] | 王根荣;陆慧.标准物质及其管理[J].化学分析计量,2004(6):59-62. |
[2] | 王巧云;何欣;王锐.国内外标准物质发展现状[J].化学试剂,2014(4):289-296. |
[3] | 方奕文;郝志峰;余林;陈善文;孙长勇;宋一兵.ICP-AES法直接测定海泥中的Fe和Mn[J].分析试验室,2003(4):87-88. |
[4] | 王干珍;汤行;叶明;严慧;易晓明;郭腊梅.电感耦合等离子体原子发射光谱法测定含碳质钒矿石中硅铝铁钒磷[J].冶金分析,2016(5):30-34. |
[5] | 杨晓燕;李莉;王碧璇;肖菊花;朱琦.用ICP-AES法测定电镀铬废水中的铬含量[J].材料保护,2008(9):70-71. |
[6] | 杨佩;王浩杰;陈晓泉;伏荣进;杨鹔;龚诚;王鑫磊.黄金掺铱制品的金相研究[J].贵金属,2014(z1):168-170. |
[7] | 王烨;颜芝.我国贵金属饰品成色检验进展[J].岩矿测试,2003(4):284-290. |
[8] | 姜岚.贵金属无损检测消除干扰因素对测试结果影响方法的研究[J].计量与测试技术,2010(09):1-3. |
[9] | 陈永红;黄蕊;陈菲菲;张雨.ICP-MS测定高纯金中的杂质元素[J].黄金,2009(4):43-47. |
[10] | 伏荣进;曲蔚;华毅超;司玮.ICP-AES法测定金箔的含金量[J].中国测试技术,2007(3):70-72. |