对电色谱塔板高度方程正确与否进行了验证.对于电色谱塔板高度方程的若干影响因素如色谱柱结构参数、毛细管内半径r0、保留因子k、扩散系数Dm以及脱附速率常数kd对柱效率的影响进行了实验考察,并将实验数据与借助计算机计算的结果加以对比.从而得出结论:我国色谱理论工作者提出的电色谱塔板高度方程成立.由于该方程合理地考虑到各种影响溶质谱带展宽的因素,因此可以为毛细管电色谱过程的柱型设计、条件优化提供理论上的指导.
参考文献
[1] | Jorgenson J W, Lukaca K D. J Chromatogr A, 1981, 218: 209 |
[2] | Tsuda T, Nomura K, Nakagawa G J. J Chromatogr A, 1982, 248: 241 |
[3] | Wen E, Asiaic R, Horvath Cs. J Chromatogr, 1999, 855: 348 |
[4] | Knox J H. Chromatographia, 1988, 26: 329 |
[5] | Cikalo M G, Bartle K D, Myers P. J Chromatogr A, 1999, 836: 35 |
[6] | Dai Chaozheng. Chinese Journal of Chromatography, 1999, 17(6): 514戴朝政. 色谱, 1999, 17(6): 514 |
[7] | Horvath Cs, Lin H J. J Chromatogr, 1976, 126: 401 |
[8] | Horvath Cs, Lin H J. J Chromatogr, 1978, 149: 43 |
[9] | Zhang Yukui, Zou Hanfa, Bao Miansheng, Dong Lifu, Lu Peizhang (Lu Peicheng). Chinese Journal of Chromatography, 1989, 7(2): 70张玉奎, 邹汉法, 包绵生, 董礼孚, 卢佩章. 色谱, 1989, 7(2): 70 |
[10] | Gusev I, Huang X, Horvath Cs. J Chromatogr A, 1999, 855: 273 |
[11] | Ericson C, Liao J L, Nakazato K, Hjerten S. J Chromatogr A, 1997, 767: 33 |
[12] | Rathore A S, Horvath Cs. Anal Chem, 1998, 70: 3271 |
[13] | Remco S, Poppe H. J Chromatogr A, 1999, 853: 45 |