建立了一种快速、灵敏的细径柱高效液相色谱测定非离子造影剂欧乃帕克的方法.用高氯酸去除血浆蛋白,流动相采用简单的7%(体积分数)甲醇水溶液体系.方法的线性范围为0.5~500 mg/L,相关系数大于0.999 8.欧乃帕克的最低检测量为154 pg,保留时间为(3.46 ± 0.02) min,总分析时间少于5 min.日内、日间相对标准偏差(RSD)分别小于4.3%和11.4%.方法的平均回收率大于91.51%.