建立了固相萃取-离子色谱(SPE-IC)测定饮用水中痕量卤代乙酸(HAAs)(包括一氯乙酸、二氯乙酸、三氯乙酸、一溴乙酸和二溴乙酸)的方法.固相萃取采用LiChrolut EN SPE柱来进行痕量待测物的预浓缩(25倍)和基体杂质的消除,用NaOH(10 mmol/L)洗脱;色谱分离采用亲水性、高容量、氢氧化物选择型阴离子交换柱Dionex IonPac AS16(250 mm×4 mm i.d.),以NaOH为流动相进行浓度梯度淋洗,淋洗速度为0.8 mL/min,电导检测,进样量为500 μL.结果表明,用SPE-IC法测定HAAs,一溴乙酸的检测限为12.5 μg/L,其余4种HAAs的检测限为0.38~1.69 μg/L.该法可实现对饮用水中痕量卤代乙酸的测定.
参考文献
[1] | Hanson M L,Solomon K R.Environmental Pollution,2004,130:371 |
[2] | Rodriguez M J,Serodes J B,Levallois P.Water Research,2004,38:4 367 |
[3] | US EPA.National Primary Drinking Water Regulation:Interim Enhanced Surface Water Treatment,1998,63(241):69 477 |
[4] | Hodgeson J W,Collins J,Barth R E.EPA Method 552.0,Revision 1.0.Cincinnati (OH,USA):National Exposure Research Laboratory,Office of Research and Development,US Environmental Protection Agency,1990 |
[5] | Munch D J,Munch J W,Pawlecki A M.EPA Method 552.2,Revision 1.0.Cincinnati (OH,USA):National Exposure Research Laboratory,Office of Research and Development,US Environmental Protection Agency,1995 |
[6] | Barron L,Paull B.J Chromatogr A,2004,1 047:205 |
[7] | Paull B,Barron L.J Chromatogr A,2004,1 046:1 |
[8] | Liu Yongjian,Mou Shifen.J Chromatogr A,2003,997:225 |
[9] | Barron L,Nesterenko P N,Paull B.J Chromatogr A,2005,1 072:207 |
[10] | Loos R,Barcelo D.J Chromatogr A,2001,938:45 |
[11] | Martinez D,Borrull F,Calull M.J Chromatogr A,1998,827:105 |
[12] | Hautman D P,Munch D J.US EPA Method 326.0,Revision 1.0.Cincinnati(OH,USA):Office of Ground Water and Drinking Water,US Environmental Protection Agency,2002 |
[13] | Slingsby R,Kiser R.TrAC,2001,20(6/7):288 |