建立了毛细管电泳-紫外检测同时测定血清中左旋多巴和甲基多巴的方法.以40 mmol/L硼砂(pH 9.5)为分离缓冲溶液,在3.45 kPa(0.5 psi)压力下进样7 s、分离电压22 kV、检测波长200 nm、温度25 ℃的条件下进行测定,两种物质获得了较好的分离.甲基多巴和左旋多巴分别为1.0~64.0 mg/L和1.0~71.0 mg/L时与峰面积呈良好的线性关系,线性相关系数分别为0.999 8和0.999 4,检出限分别为0.6和0.8 mg/L(以信噪比为3计).将该法用于血清中甲基多巴和左旋多巴的测定,回收率为82.8% ~88.8% ,相对标准偏差为2.10% ~2.63% .