利用大体积进样技术(large volume injection,LVI),结合气相色谱-质谱方法对二(噁)英的测定效果进行了研究.同时与传统分流/不分流进样技术进行了对比.对进样体积为1,5,10,25,50和100 μL的色谱图进行了分析.研究表明使用大体积进样方式,在不影响色谱分离度的同时,大幅度提高了分析灵敏度.通过对土壤样品的检测,证明该方法可以用于环境样品的实际测定.
参考文献
[1] | Lindstrm G,Hooper K,Petreas M,Stephens R,Gilman A.Environ Health Persp,1995,103(Suppl 2):135 |
[2] | UNEP.Final Act of the Conference of Plenipotentiaries on the Stockholm Convention on Persistent Organic Pollutants.UNEP/POPs/CONF/4,2001 |
[3] | Sing S B,Kulshrestha G.J Chromatogr A,1997,774:97 |
[4] | Liem A K D.Trends Anal Chem,1999,18:499 |
[5] | Norlock F M,Jang J K,Zou Q,Schoonover T M,Li A.Journal of the Air & Waste Manage Assoc,2002,52(1):19 |
[6] | Liu L.Analytical Instrumentation (刘玲.分析仪器),2005(2):16 |
[7] | Neil C,Pete S.Environmental Chemistry (环境化学),2005,24(2):223 |
[8] | Liao H R,Wu X M,Zeng H H,Xu B Z,Mai B C.Chinese Journal of Chromatography (廖华勇,吴雪梅,曾红惠,徐碧珠,麦伯成.色谱),2003,21(4):414 |
[9] | Liu H X,Zhang Q H,Cai Z W,Li A,Wang Y W,Jiang G B.Anal Chim Acta,2006,557:314 |