我们建立了一套基于高温超导量子干涉仪(SQUID)的可靠的、经济型的单通道心磁图(MCG)系统,该系统在一个简易的磁屏蔽室(MSR)中运行.实验结果表明,对于消除心磁测量中的环境噪声,简易磁屏蔽室和有源补偿相结合是一个很合适的方案,这种方案对心磁图的普及有实际的帮助.同时,本文给出了一个有源补偿的简单模型,并将计算结果与实验数据进行了比较.
We have constructed a reliable and economical single-channel magnetocardiogram (MCG) system based on high-Tc SQUIDs operated in a simply magnetically shielded room (MSR). Experimental results show that the combination of a simple MSR and an active shielding is an appropriate scheme for the environmental noise cancellation in MCG measurements and might be a practical way towards the popularization of MCGs. Meanwhile, a simple model for active shielding is presented and compared with experimental data.
参考文献
[1] | D Cohen;E A Edelsack;J E Zimmerman .[J].Applied Physics Letters,1970,16:278. |
[2] | D.Platzek;H.Nowak .Low-T_c magnetometer used for active shielding in the frequency range dc-500 Hz in biomagnetic[J].Superconductor Science & Technology,1999(11):940-942. |
[3] | M H m l inen et al.[J].Reviews of Modern Physics,1993,65:413. |
[4] | L Kaufman S;J Williamson.Recent developments in neuromagnetism[A].Butterworth,Boston,1987:100-113. |
[5] | C R Davidson.[A].,2000:35-38. |
[6] | H J M ter Brake;R Huonker;H Rogalla .[J].Measurement Science and Technology,1993,4:1370. |
[7] | G H Chen.[A].Germany,2002:563-565. |
[8] | X M Zhu;Y Tian;S P Zhao .A noise feedback least-mean-square algorithm of data processing for SQUID-based magnetocardiography[J].Superconductor Science & Technology,2005(8):1054-1059. |
[9] | S N Erné.[A].,1980:51-77. |
[10] | H Kaden.Wirbelstr? me und Schirmung in der Nachrichtentechnik.2.Ed[J].Journal of F Bergmann Mtünchen,1959:1-354. |
[11] | K J Best;J Bork .[J].ETZ,1989,110(16):814. |
[12] | E Baumand J;Bork J Magn .[J].Journal of Magnetism and Magnetic Materials,1991,101:69-74. |
[13] | M Chen;R T Wakai.[A].Germany,2002:653-654. |