基于微乳液有增稳、增敏、增溶等作用,建立了微乳液-鲁米诺-H2O2-Fe(Ⅱ)化学发光体系测定Fe(Ⅱ)的分析方法.该法测定Fe(Ⅱ)的浓度范围为1×10-6~1×10-4 g/L,检出限为1.20×10-7 g/L.对浓度为1×10-6 g/L的Fe(Ⅱ)标准溶液连续平行测定11次,相对标准偏差(RSD)为1.33%.利用EDTA为掩蔽剂,可有效消除大多数元素离子对测定Fe(Ⅱ)的干扰,从而可用于矿样中Fe(Ⅱ)的测定,回收率在96%~107%之间.

参考文献

[1] Qin Wei;Bin Du .Rapid and selective method for the spectrophotometric determination of nickel naphthenate in gasoline in a microemulsion[J].Talanta: The International Journal of Pure and Applied Analytical Chemistry,1998(5):957-961.
[2] Bin Du;Qin Wei;Shuren Wang;Weile Yu .Application of microemulsions in determination of chromium naphthenate in gasoline by flame atomic absorption spectroscopy[J].Talanta: The International Journal of Pure and Applied Analytical Chemistry,1997(10):1803-1806.
[3] Qin Wei;Jinhu Yang;Yong Zhang;Guohua Chang;Bin Du .Determination of antimony(III) in environmental water samples in microemulsion system by the fluorescence quenching method[J].Talanta: The International Journal of Pure and Applied Analytical Chemistry,2002(3):419-426.
[4] 李绍卿,杨育青,朱亚鹏.鲁米诺-H2O2-Fe(Ⅲ)-Tween-80化学发光与应用[J].分析测试技术与仪器,2002(01):53-55.
[5] 吴秋华,李正平,方正.鲁米诺化学发光体系应用新进展[J].化学试剂,2005(04):212-216.
[6] 张丽丽,臧利国,陈蓁蓁,唐波.化学发光分析法应用进展[J].理化检验-化学分册,2004(04):243-248.
[7] 季宁宁,徐法君,刘道杰.鲁米诺的应用及进展[J].聊城大学学报(自然科学版),2003(04):31-32,59.
[8] 王莉平,李绍卿.化学发光法测定铜的进展[J].长安大学学报(建筑与环境科学版),2003(04):75-78.