通过对氧化锑试样分解方法、硅磷分析线选择、共存元素光谱干扰和仪器最佳分析条件等方面的研究,建立了电感耦合等离子体原子发射光谱法测定氧化锑中硅和磷的方法.选用谱线Si 251.611 nm,P 214.914 nm作分析线,避免了基体元素对硅、磷的干扰.在优化的实验条件下,当硅磷的含量在0.01~300 mg/L范围内时,其吸先度与硅磷浓度呈良好的线性关系,方法的检出限为Si 0.04 mg/L,P 0.07 mg/L.对4个样品的硅和磷进行测定,回收率分别为95.82%~107.12%和93.47%~98.52%,相对标准偏差(n=10)分别为0.6%~3.4%和0.8%~4.8%.
参考文献
[1] | 傅明,胡宇东,杨万彪,陈新焕.火焰原子吸收法连续测定氧化锑中铅、铜、铁含量[J].光谱学与光谱分析,2003(05):997-998. |
[2] | 鲁静冬,董永平,顾晓川.磷钼蓝光度法测定钢中磷[J].冶金分析,2004(04):77-78. |
[3] | 汤成兰.钼蓝光度法快速测定铬铁中硅[J].冶金分析,2006(06):92-92. |
[4] | Xu YY;Zhou HF;Wang GX;Zhou JF;Tao GH .Determination of trace amounts of lead, arsenic, nickel and cobalt in high-purity iron oxide pigment by inductively coupled plasma atomic emission spectrometry after iron matrix[J].Analytica chimica acta,2007(1):204-209. |
[5] | 莫庆军,宋兰会,王秀艳.ICP-AES法测定锰铁中硅和磷方法研究[J].光谱学与光谱分析,2004(12):1666-1668. |
[6] | Yang CH;Jiang SJ .Determination of B, Si, P and S in steels by inductively coupled plasma quadrupole mass spectrometry with dynamic reaction cell[J].Spectrochimica Acta, Part B. Atomic Spectroscopy,2004(9):1389-1394. |
[7] | 张兰,关雄俊,莫达松.ICP-AES法测定三氧化二锑中杂质[J].光谱实验室,2001(04):520-522. |
[8] | 袁齐.ICP-AES直接测定锑产品中杂质元素[J].光谱学与光谱分析,2001(04):540-541. |