欢迎登录材料期刊网

材料期刊网

高级检索

为检验新型瓣膜的疲劳磨损特性,采用 TH-2200型人工心瓣疲劳寿命测试仪对聚甲醛瓣叶的双叶心脏瓣膜进行加速疲劳实验,并对疲劳后的瓣膜进行扫描电镜检测,发现该瓣膜经4×108次疲劳测试后,瓣环无明显磨损,瓣叶转动灵活,心瓣正常工作,但在瓣环枢轴坑的部分位置发现少量粘附的聚甲醛颗粒,尺寸约0.1~1μm,瓣叶总磨损率为0.681%.虽然新型瓣膜疲劳实验证明了瓣膜结构设计合理,能够满足国家标准要求的疲劳寿命,但是仍需要对瓣叶材料表面进行处理,进一步提高瓣叶的耐磨性.

In order to test the fatigue wear characteristics of the new valve,TH-2200 fatigue life tester was used for an artificial bileaflet heart valves with polyformaldehyde (POM)leaflets to have accelerated fatigue tests, surface topography was detected by scanning electron microscopy (SEM).After the 4×10 8 times fatigue tests, it was shown that no significant wear on the valve ring,valve leaflets rotated flexibility and the heart valves worked well,but a small POM pit adhesion was found in the valve ring part of the pivot position,POM parti-cles size was about 0.1-1 μm,leaflets total wear rate was 0.681%.Although the new valve fatigue test showed that the valve had reasonable structural design,which could meet the requirements of national standard fatigue life,but the valve still need to surface treatment,to improve the wear resistance of the valve leaflets.

参考文献

[1] Mohammadi H;Mequanint K .Prosthetic aortic heart valves: modeling and design.[J].Medical engineering and physics,2011(2):131-147.
[2] Kutting M;Roggenkamp J;Urban U;Schmitz Rode T;Steinseifer U .Polyurethane heart valves: past, present and future.[J].Expert review of medical devices,2011(2):227-233.
[3] Rahimtoola SH .Choice of prosthetic heart valve in adults an update.[J].Journal of the American College of Cardiology,2010(22):2413-2426.
[4] Philippe Pibarot;Jean G. Dumesnil .Prosthetic Heart Valves Selection of the Optimal Prosthesis and Long-Term Management[J].Circulation: An Official Journal of the American Heart Association,2009(7):1034-1048.
[5] Shahriari,S.;Maleki,H.;Hassan,I.;Kadem,L. .Evaluation of shear stress accumulation on blood components in normal and dysfunctional bileaflet mechanical heart valves using smoothed particle hydrodynamics[J].Journal of Biomechanics,2012(15):2637-2644.
[6] Haggag Y M .The central axis prosthetic cardiac valve:an in vitro study of pressure drop assessment under steady-state flow conditions[J].Journal of Biomedical Engineering,1990,12(1):63-68.
[7] Grigioni M.;D'Avenio G.;Calcagnini G.;Barbaro V.;Daniele C. .Wear patterns from fatigue life testing of a prosthetic heart valve[J].Wear: an International Journal on the Science and Technology of Friction, Lubrication and Wear,1999(2):743-748.
[8] Jozwik K;Karczemska A .The new generation Ti6Al4V artificial heart valve with nanocrystalline diamond coat-ing on the ring and with derlin disc after long-term me-chanical fatigue examination[J].Diamond and Related Materials,2007,16(4-7):1004-1009.
[9] 刁颖敏;席葆树.人工心脏主动脉瓣膜的定常流实验研究[J].北京生物医学工程,1982(01):23-28.
[10] 褚银平,程金莲,刘爱军,高奇英,张羽,张栋.脉动流下机械心脏瓣膜瓣阀开启状态的评价☆[J].中国组织工程研究,2012(40):7480-7485.
[11] Zilla P;Brink J;Human P;Bezuidenhout D .Prosthetic heart valves: Catering for the few[J].Biomaterials,2008(4):385-406.
[12] 龙春光,吴茵,华熳煜.聚甲醛多元复合材料的力学性能和承载能力[J].功能材料,2005(05):684-686.
[13] Jinyao Chen;Ya Cao;Huilin Li .Investigation of the friction and wear behaviors of polyoxymethylene/linear low-density polyethylene/ethylene-acrylic-acid blends[J].Wear: an International Journal on the Science and Technology of Friction, Lubrication and Wear,2006(11/12):1342-1348.