简单介绍了室温磁热效应测量仪的组成、工作原理及其优点;用金属Gd作为标准样品,用两个不同尺寸的温度传感器,测量了两个大小不同的样品.结果表明,相同条件下测量的数据重复性很好,而不同条件下测量的数据有一些差别,对此进行了分析,发现测量样品时特别是样品较小时,温度传感器的热容以及绝热条件等因素是引起误差的主要原因.
参考文献
[1] | Pecharsky VK;Gschneidner KA .Giant magnetocaloric effect in Gd-5(Si2Ge2)[J].Physical review letters,1997(23):4494-4497. |
[2] | V. K. Pecharsky;K. A. Gschneidner;Jr. .Tunable magnetic regenerator alloys with a giant magnetocaloric effect for magnetic refrigeration from ~20 to ~290 K[J].Applied physics letters,1997(21/24):3299-3301. |
[3] | Tegus O;Bruck E;Buschow K H J et al.Transition-metal-based magnetic refrigerants for room-temperature appllcations[J].Nature,2002,415:150-152. |
[4] | Pecharsky V K;Gschneidner K A .Comparison of the magnetocaloric effect derived from heat capacity,direct,and magnetization measurements[J].Advances in Cryogenic Engineering,1996,42:425-430. |
[5] | Karl A. Gschneidner Jr.;Vitalij K. Pecharsky .The influence of magnetic field on the thermal properties of solids[J].Materials Science & Engineering, A. Structural Materials: Properties, Misrostructure and Processing,2000(2):301-310. |
[6] | 黄焦宏,王正德,金培育,刘金荣,张泽玉,王贵,杨保荣,成永顺,邱巨峰,徐来自.Gd5(SixGe1-x)4合金磁热效应的直接测量[J].包头钢铁学院学报,2000(03):212-214. |
上一张
下一张
上一张
下一张
计量
- 下载量()
- 访问量()
文章评分
- 您的评分:
-
10%
-
20%
-