经测试原理分析,证明了高温空腔(测试腔体内无样品)特性(主要指谐振频率和品质因数)的准确性和稳定性是保证样品高温介电性能测试结果准确性和稳定性的必要条件.进行了室温~1 600℃定点温度下恒温时间为3、5和8 min、7~18 GHz的空腔特性测试,结果表明谐振频率相互之间相对偏差±0.05%,品质因数相互之间相对偏差<±5%,验证了恒温时间为3 min时可获得准确的高温空腔特性数据.还在不同时间不同季节多次对空腔进行了室温~1 600℃、7~18 GHz空腔特性测试,结果表明谐振频率之间相对误差<±0.04%,品质因数之间的相对误差<±5%,进一步验证了该测试系统可获得准确稳定的测试结果.
参考文献
[1] | Westphal W B .Dielectric constant and loss data[MA 02139,AD A121008][R].laboratory for Insulation research massachusetts institute of technology cambridge |
[2] | Westphal W B;Iglesias J .Dielectric measurements on high-temperature materials.[AD873038][R].Air Force Materials Laboratory |
[3] | Gilreath M C.High-temperature dielectric properties of candidate space-shuttle thermal-proection-system and antennawindow materials[M].NASA Langley Research Center,1974:1-52. |
[4] | Anderson B L.Highly parallel optical true time delay device for phased array antennas[A].Monteney,CA,1996 |
[5] | ASTM D2520-1995.Standard test methods for complex permittivity(dielectric constant)of solid electrical Insulating materials at microwave frequencies and temperatures to 1 650℃[S]. |