欢迎登录材料期刊网

材料期刊网

高级检索

首先对有限元软件Ansys中镍钛合金材料模型的准确性进行了研究,然后对两种镍钛合金支架的抗压缩性能进行了模拟分析与实验验证,结果证明有限元法可以在一定程度上替代支架原型的测试工作.在此基础上,又进一步系统地分析了一种支架设计中筋的尺寸变化对支架抗压缩性能的影响.结果显示,增加支架筋的宽度和厚度都能够提高支架的抗压缩性能,且这两个方向上尺寸的增加对提高支架抗压缩性能的作用基本相当.

参考文献

[1] Andrew J Carter;Douglas Scott;John R Laird et al.Progressive Vascular Remodeling and Reduced Neointimal Formation After Placement of a Thermoelastic Self-Expanding Nitinol Stent in an Experimental Model[J].Catheterization and Cardiovascular Interventions,1998,44:193-201.
[2] Yu ZX;Tamai H;Kyo E;Kosuga K;Hata T;Okada M;Nakamura T;Komori H;Tsuji T;Takeda S;Motohara S;Uehata H .Comparison of the self-expanding Radius stent and the balloon-expandable Multilink stent for elective treatment of coronary stenoses: a serial analysis by intravascular ultrasound.[J].Catheterization and cardiovascular interventions: Official journal of the Society for Cardiac Angiography & Interventions,2002(1):40-45.
[3] Ted Feldman .Unique Properties of Self-Expanding Stents[J].Catheterization and Cardiovascular Interventions,2000,49:408-409.
[4] Tan L B;Webb D C;Kormi K et al.A method for investigating the mechanical properties of intracoronary stents using finite element numerical simulation[J].International Journal of Cardiology,2001,78:51-67.
[5] Migliavacca F;Petrini L;Colombo M;Auricchio F;Pietrabissa R .Mechanical behavior of coronary stents investigated through the finite element method.[J].Journal of Biomechanics,2002(6):803-811.
[6] Etave F;Finet G;Boivin M;Boyer JC;Rioufol G;Thollet G .Mechanical properties of coronary stents determined by using finite element analysis.[J].Journal of Biomechanics,2001(8):1065-1075.
[7] 杨大智;吴明雄.Ni-Ti形状记忆合金在生物医学领域的应用[M].北京:冶金工业出版社,2003:9.
[8] Whitcher F D .Simulation of in vivo loading conditions of Nitinol vascular stent structures[J].Computers and Structures,1997,64:1005-1011.
[9] Petrini L;Migliavacca F;Dubini G.Numerical analysis of vascular stents exploiting shape-memory-alloy behavior[A].Italy,2003
[10] Rieu R;Barragan P;Masson C;Fuseri J;Garitey V;Silvestri M;Roquebert P;Sainsous J .Radial force of coronary stents: a comparative analysis.[J].Catheterization and cardiovascular interventions: Official journal of the Society for Cardiac Angiography & Interventions,1999(3):380-391.
[11] Schrader CS;Beyar R .Evaluation of the compressive mechanical properties of endoluminal metal stents[J].Catheterization and Cardiovascular Interventions,1998,44:179-187.
[12] Moore J E;Joel L Berry .Fluid and solid mechanical implications of vascular stenting[J].Annals of Biomedical Engineering,2002,30:498-502.
[13] Frank A O;Walsh P W;Moore J E .Computational fluid dynamics and stent design[J].Artificial Organs,2002,26:614-625.