欢迎登录材料期刊网

材料期刊网

高级检索

通过L-S相转化法制备了膜式氧合器的聚砜不对称膜,并对其性能进行表征.制膜实验结果表明,随着铸膜液浓度的增加,氧气和二氧化碳渗透通量减小,临界透水压力增大,而湿膜的预挥发时间越长,膜的通量越小,但是临界透水压力变化不大,因此优化的铸膜液浓度为15%和预挥发时间为5s.此外针对膜式氧合器的气液传输测试结果表明,当牛血流速为5 L/min时,氧气和二氧化碳通过膜的传榆速率分别达到163 mL/min和148 mL/min,可以满足膜式氧合器气体交换传输能力的要求.

参考文献

[1] Lim M W .The history of extracorporeal oxygenator[J].ANAESTHESIA,2006,61:984-995.
[2] Zwischenberger BA;Clemson LA;Zwischenberger JB .Artificial lung: progress and prototypes.[J].Expert review of medical devices,2006(4):485-497.
[3] Tatsumi E .Artificial lungs: current state and trends of clinical use and research and development.[J].Journal of artificial organs: The official journal of the Japanese Society for Artificial Organs,2007(1):1-5.
[4] Frank wiese.Membrane for artificial lung[A].:49-68.
[5] Gray D N.Polymeric membrane for artificial lung[M].American chemical society:Washington DC,1984:150-161.
[6] Don N Hiroyoshi Kawakami .Polymeric membrane materials for artificial organs[J].Artif Orgasns,2008,11:177-182.
[7] Don N Tadashi Motomura .Tomohiro Maeda Development of silicone rubber hollow fiber membrane oxygenator for ECMO[J].Artificial Organs,27(11):1050-1056.
[8] Mário DRUMMOND;Domingo M .BRAILE Technological evolution of membrane oxygenators[J].Braz J Cardiovasc Surg,2005,20(04):432-437.
[9] 杨瑜静,谷雪莲,徐秀林,朱根娣,王新华.复用血液透析器(人工肾)膜表面粘附蛋白的实验研究[J].膜科学与技术,2008(06):79-83,98.
[10] Wang DL.;Teo WK.;Li K. .Preparation and characterization of high-flux polysulfone hollow fibre gas separation membranes[J].Journal of Membrane Science,2002(1/2):247-256.
[11] Robles N R;Murga L;Galvan S;Esparrago J F et al.Hemodialysis with cuprophane or polysulfone:effects on uremic polyneuropathy[J].Am J disc,1993,21(03):282-287.
[12] Fawen Wu;Lei Li;Zhihong Xu;Shujuan Tan;Zhibing Zhang .Transport study of pure and mixed gases through PDMS membrane[J].Chemical engineering journal,2006(1):51-59.
[13] Machado P.S.T.;Borges C.P.;Habert A.C. .Membrane formation mechanism based on precipitation kinetics and membrane morphology: Flat and hollow fiber polysulfone membranes[J].Journal of Membrane Science,1999(2):171-183.
[14] Motohiro Niwa;Hiroyoshi Kawakami;Shoji Nagaoka .Development of a novel polyimide hollow-fiber oxygenator[J].Artificial Organs,28(05):487-495.
[15] Wiliam J;Federspiel Kristie A;Henchir Lung.Artificial:Basic Principles and Current Apphcations[M].University of Pittsbrngh,Pennsylvania,U.S.A:909-921.
[16] Visser C;Jong D S D .Clinical evaluation of six hollow-fibre membrane oxygenators[J].Univers Hosp Maastr perf,1997,12:357-368.
上一张 下一张
上一张 下一张
计量
  • 下载量()
  • 访问量()
文章评分
  • 您的评分:
  • 1
    0%
  • 2
    0%
  • 3
    0%
  • 4
    0%
  • 5