介绍了用于中子探测器效率刻度的薄壁快脉冲~(252)Cf裂变电离室.在用飞行时间法测量中子能谱时,该裂变电离室能够给出~(252)Cf裂变中子发射时刻信号,输出脉冲上升时间约为5.5 ns,电离室由厚度为0.15 mm的不锈钢构成.测试结果表明,对裂变碎片的探测效率为99.2%,α粒子脉冲幅度和碎片脉冲幅度可清晰分开.

A thin wall, fast pulse ~(252)Cf ionization chamber, which was designed for calibrating efficiency of neutron detectors, was described. The ionization chamber can be used as the start detector in time-of-flight measurements of the ~(252)Cf fission neutron spectrum. The chamber is composed of 0.15 mm thick stainless steel, the rise time of pulse signals is about 5.5 ns. As the test result shows, the detection efficiency of fission fragments is 99.2%, and pulse signals caused by fissions are distinguished clearly from those caused by α decays.

参考文献

[1] Mannhart W.IAEN-INDC(NDS)-220,1989,305-336.
[2] Chen Jinxiang,Shi Zhaomin,Tang Guoyou.Nuclear Electronics and Detection Technology,1993,13(6):323(in Chinese).(陈金象,施兆明,唐国有.核电子学与探测技术,1993,13(6):323.)
[3] Cub J,Finekh E,Gebhardt K,et al.Nucl Instr and Meth 1989,A274:217.
[4] Li Anli,Wang Xiaozhong,Bai Xixiang,et al.Atomic Energy Science and Technology,1994,28(2):132(in Chinese).(李安利,王效忠,白希祥等.原子能科学技术,1994,28(2):132.)
[5] Li Jiansheng,Zhang Yi,Jin Yu,et al.Nuclear Electronics and Detection Technology,2001,21(4):264(in Chinese).(李建胜,张翼,金宇等.核电子学与探测技术,2001,21(4):264.)
[6] Meng Jiangchen,Huang Shengnian.Nuclear Electronics and Detection Technology,1989,9(1):1(in Chinese).(孟江辰,黄胜年.核电子学与探测技术,1989,9(1):1.)
[7] Bao Shanglian,Liu Wenlong,Fan Tieshuan,el at.Atomic Energy Science and Technology,2001,35(5);41(in Chinese).(包尚联,刘文龙,樊铁栓等.原子能科学技术,2001,35(5):41.)
[8] Chalupka A.Nucl Instr and Meth,1979,164:105.